[Water vapor permeability measurement of the materials used as a wound covering]
1Trauma Center Postgraduate Medical College and 304th Hospital of PLA, Beijing 100037.
Abstract:
This paper introduces a new apparatus and method that suits to measuring the water vapor transmission rate (WVTR) of wound covering materials. WVTR of synthetic materials, amniotic membrane, porcine and cadaveric skin was measured under different conditions of four temperatures, three mediums, and "water cup" or "inverted water cup". The temperature was in close relationship with the WVTR except Omiderm, amniotic membrane, and cadaveric skin between 30 degrees C and 37 degrees C groups. The medium did not affect the WVTR (P > 0.05). There was significant difference between the "water cup" and "inverted water cup" groups(P < 0.01). This study suggests that the apparatus is easy-to-operate, reliable and suitable for the measurement of WVTR of burn wound coverings and other membranous materials.
Related Concept Videos
Permeability of Concrete
Vapor Pressure of Fluid
When a liquid is placed in a closed container with a small air space, and the space is evacuated, vapor molecules will...
Surface Tension, Capillary Action, and Viscosity
The various IMFs between identical molecules of a substance are examples of cohesive forces. The molecules within a liquid are surrounded by other molecules and are attracted equally in all directions by the cohesive forces within the liquid. However, the molecules on the surface of a liquid are attracted only by about one-half as many molecules. Because of the unbalanced molecular attractions on the surface molecules, liquids contract to form a shape that minimizes the number...