Closed-loop controlled inspired oxygen concentration for mechanically ventilated very low birth weight infants with
N Claure1, T Gerhardt, R Everett
1Division of Neonatology, Department of Pediatrics, University of Miami School of Medicine, Miami, Florida 33101, USA. nclaure@miami.edu
Background:
Mechanically ventilated very low birth weight infants often present with frequent episodes of hypoxemia, and maintaining arterial oxygen saturation by pulse oximetry (SpO(2)) within a normal range by manual fraction of inspired oxygen (FIO(2)) adjustments is difficult and time consuming.
Objectives:
An algorithm for closed-loop FIO(2) control (cFIO(2)) to maintain SpO(2) within a target range was compared with continuous manual FIO(2) (mFIO(2)) adjustments by a nurse in a group of ventilated infants who presented with frequent episodes of hypoxemia.
Results:
Fourteen infants (birth weight: 712 +/- 142 g; gestational age: 25 +/- 1.6 weeks; age: 26 +/- 11 days; synchronized intermittent mandatory ventilation rate: 24 +/- 10 b/m; peak inspiratory pressure: 17.5 +/- 2.0 cmH(2)O; positive end-expiratory pressure: 4.3 +/- 0.5 cmH(2)O) were studied for 2 hours on each mode in random sequence. Both modes aimed to maintain SpO(2) between 88% and 96%. There were 15 +/- 7 and 16 +/- 6 hypoxemic episodes/hour (SpO(2) <88%, >5 s) during mFIO(2) and cFIO(2), respectively; episode duration was 41 +/- 23 and 32 +/- 15 s, totaling 19 +/- 16% and 17 +/- 12% of recording time. There were 13 +/- 10 and 10 +/- 8 hyperoxemic episodes/hour (SpO(2)>96%, >5 s) during mFIO(2) and cFIO(2,) respectively; episode duration was 27 +/- 15 and 24 +/- 19 s, totaling 15 +/- 14% and 10 +/- 9% of recording time. Mean SpO(2) and FIO(2) levels were similar during both modes. The nurse made 29 +/- 17 adjustments/hour during mFIO(2). There was a significant increase in the duration of normoxemia (SpO(2) between 88%-96%) during cFIO(2) (75 +/- 13 vs 66 +/- 14% of recording time).
Conclusion:
In this group of infants, cFIO(2) was at least as effective as a fully dedicated nurse in maintaining SpO(2) within the target range, and it may be more effective than a nurse working under routine conditions. We speculate that during long-term use, cFIO(2) may save nursing time and reduce the risks of morbidity associated with supplemental oxygen and episodes of hypo- and hyperoxemia.
More Related Videos
05:45Delivery of In Vivo Acute Intermittent Hypoxia in Neonatal Rodents to Prime Subventricular Zone-derived Neural Progenitor Cell Cultures
Published on: November 2, 2015
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Related Concept Videos
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)