Related Experiment Video
Updated: Jul 30, 2026

A New Approach for the Comparative Analysis of Multiprotein Complexes Based on 15N Metabolic Labeling and Quantitative Mass Spectrometry
Published on: March 13, 2014
Improved method for the measurement of large neutral amino acids in biological matrices
R Bongiovanni1, B K Yamamoto, G E Jaskiw
1Veterans Affairs Medical Center, GMH(B), Brecksville, OH 44141, USA. rbongi4333@aol.com
Abstract:
A high-performance liquid chromatographic method for measuring neutral amino acids in rat sera, brain tissues, and perfusates was developed by using o-phthalaldehyde sulfite as a pre-column derivatization reagent. With the present method, it was possible to separate the neutral amino acids within a single run in 25 min, while the acidic amino acids were eluted near or at the solvent front. The recovery was above 88.8% with a relative standard deviation (RSD) below 4.2%. The within- and between-day assay reproducibility for the determination of rat serum amino acids showed RSDs below 1.35 and 7.61%, respectively. In the present study, the neutral amino acids were assayed with high sensitivity, accuracy and good reproducibility in a relatively short time and on a small sample size.