ICD malfunction caused by excessive defibrillation charge time.

A de Meester1, P H Gilles

  • 1Cardiology Department, Jolimont Hospital, 7100 Haine Saint Paul, Belgium.

Pacing and Clinical Electrophysiology : PACE
|May 9, 2001
PubMed
Summary

Delayed therapy from an implantable cardioverter-defibrillator (ICD) occurred due to excessive charge times. This rare malfunction necessitated generator replacement, highlighting charge time as a critical factor in ICD performance.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Successful treatment of a huge intra-mural haematoma of the ascending aorta by percutaneous coronary intervention.

Acta cardiologica·2020
Same author

[Implementation of reperfusion therapy in ST segment elevation myocardial infarction (STEMI). A policy statement from the Belgian Society of Cardiology and its working group of acute cardiology and interventional cardiology].

Revue medicale de Bruxelles·2010
Same author

[Implementation of reperfusion therapy in ST segment elevation myocardial infarction (STEMI). A policy statement from the Belgian Society of Cardiology and its working group of acute cardiology and interventional cardiology].

Revue medicale de Liege·2010
Same author

Co-administration of ezetimibe and simvastatin in acute myocardial infarction.

European journal of clinical investigation·2007
Same author

[Doppler ultrasound and aortic regurgitation. Evaluation of the sensitivity and specificity of Doppler findings].

Journal de radiologie·2004
Same author

[Recommendations for management of acute chest pain. Report of the Belgium Interdisciplinary Working Group on Acute Cardiology].

Revue medicale de Liege·2002

Area of Science:

  • Cardiology
  • Biomedical Engineering

Background:

  • Implantable cardioverter-defibrillators (ICDs) are crucial for managing life-threatening ventricular arrhythmias.
  • Patients with a history of myocardial infarction and ventricular tachycardia often receive ICDs for primary or secondary prevention.

Observation:

  • A 71-year-old male patient experienced progressively increasing defibrillation charge times on his ICD.
  • The patient collapsed during a fast ventricular tachycardia episode, with telemetric interrogation revealing an excessive ICD charge time.

Findings:

  • The prolonged ICD charge time resulted in delayed defibrillation therapy.
  • This malfunction rendered the ICD potentially ineffective, leading to a recommendation for generator exchange.

Implications:

Related Experiment Videos

  • Excessive charge time is an underrecognized cause of ICD malfunction.
  • Regular monitoring of ICD charge times is essential for ensuring device efficacy and patient safety.
  • This case underscores the need for prompt generator replacement when such anomalies are detected.