Selection of pulse oximetry equipment for ambulatory monitoring
C W Barratt1, H Vyas, B R Hayes-Gill
1School of Electrical and Electronic Engineering, University of Nottingham, University Park, UK. eexcb@hermes.nottingham.ac.uk
Primary Objective:
This communication describes the initial stage of a research project concerning the monitoring of SpO2 in infants prone to periods of spontaneous oxygen desaturation whilst freely moving around their home environment. The primary aim was to determine an appropriate probe type and site together with an assessment of the suitability of two commercially available oximeter units.
Research Design:
The study comprised 19 comparative tests, totalling 162 hours of recordings at resolution one sample every four seconds. Comparisons are drawn between probes, probe sites and pulse oximeters. MAIN OUTCOMES/RESULTS: The bias and precision is presented with respect to the probe and measurement site. Also, correlation between the trial and reference recordings is considered.
Conclusions:
It is concluded that ambulatory recording of SpO2 in infants utilizing equipment suitable for home monitoring can produce diagnostic data equivalent to that of the Ohmeda 3700 biox, but that an indication of movement artefact may be required for confirmation of accuracy. It became apparent that 'wrap around' probes, used on the index finger or big toe are the most suitable.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...