Related Experiment Video
Updated: Jul 29, 2026

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Ablation of atrial fibrillation in the rapid pacing canine model using a multi-electrode loop catheter
B Avitall1, A Urbonas, S Millard
1University of Illinois at Chicago, Department of Medicine, 60612, USA. Bavitall@uic.edu
Objectives:
This investigation details our experience using a loop catheter to ablate atrial fibrillation (AF) in dogs.
Background:
Atrial fibrillation is the most common arrhythmia and has significant morbidity. Maintenance of normal sinus rhythm (NSR) after conversion in many patients is still a challenge.
Methods:
A multi-electrode loop catheter was used to create linear atrial lesions to ablate AF in a rapid atrial pacing model in 29 dogs. Rhythm status was assessed over a six-month recovery period, after which tissue analysis was performed.
Results:
Acute conversion to NSR or atrial tachycardia (AT) was achieved in 90% of cases. Six of 26 conversions occurred after only left atrial (LA) lesions, and two after just right atrial lesions. Sixteen (62%) of 26 lesions that resulted in AF conversion were in the LA, and 11 of these 16 conversions occurred during a lesion connecting the mitral ring to the pulmonary veins. Acute conversion rate was similar with ring and coil electrodes, but AT was more frequent with coil electrodes (63% vs. 31%). At six months 80% of dogs were in NSR, 14% were in AT, and 7% remained in AF. There was an average reduction in P-wave amplitude of 64 +/- 26% after power application. Tissue analysis revealed transmural contiguous lesions when final outcome was NSR, and nontransmural/noncontiguous lesions where AF persisted.
Conclusions:
Multi-electrode loop catheters can create contiguous transmural lesions in either atrium to safely and effectively ablate AF and provide a stable long-term rhythm outcome in this dog model. The left atrium appears to be the dominant chamber that sustains AF. Atrial tachycardia is a frequent acute outcome with coil electrodes.