Validation test for climate control on air-loss supports
S I Reger1, T C Adams, J A Maklebust
1Department of Physical Medicine and Rehabilitation, Cleveland Clinic Foundation, Cleveland, OH, USA.
Objective:
To develop a simple, reproducible validation test protocol for classification of air-loss support systems.
Design:
Simultaneous experimental measurement of moisture loss and temperature reduction at the air-loss support surface-human body equivalent interface from a sweating human skin analogue.
Setting:
A hospital department of physical medicine and rehabilitation.
Other Participants:
These 3 manufacturers contributed 14 support surfaces.
Interventions:
Test support surfaces and a standard foam mattress were placed on a hospital bed. Water was circulated to a loading gauge, placed on a dry moisture reservoir, and connected to a water bath to keep the interface at 37 degrees +/- 0.5 degrees C. The loading gauge and support surface was adjusted 23cm below the water bath level and the air flow through the interface initiated. After the dry moisture reservoir came to temperature equilibrium for 30 minutes, it was replaced with a wet one that was saturated with 36g of saline. The temperature change and evaporation rate were recorded throughout a 90-minute test period.
Main Outcome Measures:
Temperature of support surface interface and evaporation rate.
Results:
Clustered data from temperature reduction and standardized rate of moisture loss yielded 3 groups of support surfaces in categories of no air loss (control), low air loss (LAL), and high air loss. The mean values of the characteristic temperature reduction and rate of moisture loss differed significantly between the groups. By multiple comparisons with Bonferroni's adjustment, the group means differed significantly for average temperature reduction (p <.017) and for standardized rate of moisture loss (p =.0001). The measured temperature change at any instant of time reflected the effect of evaporation and the opposing effect of thermal conductivity.
Conclusion:
Measurements of support interface climate change allowed for selective grouping of LAL surfaces according to rate of moisture evaporation and the resulting temperature reduction. Neither temperature change nor evaporation rate alone was sufficient to determine the microclimate characteristics of the support surface. Combined, these characteristics can effectively describe the performance of any LAL support system and may be used to define standards of performance.
More Related Videos
Related Concept Videos
Heating and Cooling Curves
For instance, the addition of heat raises the temperature of a solid; the amount of heat absorbed depends on the heat capacity of the solid (q = mcsolidΔT). According to thermochemistry, the relation between the amount of heat absorbed or released by a substance, q, and its...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...
Insulation Coordination
Masonry in Cold and Hot Weather Conditions
Other key practices include keeping masonry units and sand dry and...
Non-destructive Tests for Concrete Strength