Related Experiment Videos
Efficient otoacoustic emission protocols employed in a hospital-based neonatal screening program
S Hatzopoulos1, G Pelosi, J Petruccelli
1Department of Audiology and Center of Bioacoustics, University of Ferrara, Italy.
Insights
This study compared two newborn hearing screening methods: transient-evoked otoacoustic emissions (TEOAE) and distortion-product otoacoustic emissions (DPOAE). The distortion-product otoacoustic emissions protocol demonstrated comparable performance to the transient-evoked otoacoustic emissions protocol in identifying hearing function in newborns.
Area of Science:
- Audiology
- Neonatal Care
- Otoacoustic Emissions
Background:
- Newborn hearing screening is crucial for early detection of hearing loss.
- Transient-evoked otoacoustic emissions (TEOAE) are a standard screening method.
- Distortion-product otoacoustic emissions (DPOAE) offer an alternative protocol for hearing assessment.
Purpose of the Study:
- To evaluate the clinical performance of a DPOAE protocol for newborn hearing screening.
- To compare the efficacy of DPOAE with a standard TEOAE protocol in a large cohort.
- To assess the potential of DPOAE in a hospital-based neonatal hearing screening program.
Main Methods:
- A cohort of 250 healthy newborns was screened on the second day of life.
- Transient-evoked otoacoustic emissions (TEOAE) were recorded using click stimuli (70-75 dB SPL).
- Distortion-product otoacoustic emissions (DPOAE) were measured using an asymmetrical protocol (75-65 dB SPL) at five frequencies (1.5–5.0 kHz).
Main Results:
- The DPOAE protocol achieved a high pass rate, showing 0.98 similarity to the TEOAE protocol.
- Both TEOAE and DPOAE methods indicated normal cochlear function in the screened neonates.
- The results suggest DPOAE is a viable option for neonatal hearing screening.
Conclusions:
- The DPOAE protocol shows significant clinical potential as a reliable method for newborn hearing screening.
- A DPOAE evaluation at five pre-selected frequencies yields results comparable to TEOAE.
- This study supports the integration of DPOAE into routine neonatal hearing screening programs.
Abstract:
Within the context of a hospital-based newborn hearing screening program, we have studied the application of two OAE protocols (TEOAE and DPOAE) on a group of 250 well babies. The main goal of this study was to evaluate the performance of DPOAE protocol in a relatively large population sample. using a preset number of five tested frequencies, in comparison with a default TEOAE screening protocol. The data were collected on the second day of life and during spontaneous sleep. The TEOAE recordings were acquired with linear protocols using click stimuli of 70-75 dB SPL and were used as indicators of normal cochlear function. The cubic distortion product DPOAE responses were evoked by an asymmetrical 75-65 dB SPL protocol, with a frequency ratio of 1.22. Five frequencies (referring to F2) were tested at 1.5, 2.0, 3.0, 4.0 and 5.0 kHz. The data from the DPOAE responses show a similar pass rate (similarity = 0.98) to the linear TEOAE protocol. The data presented suggest that a DPOAE cochlear evaluation, at 5 pre-selected frequencies, has clinical potential.