Optimising miller cuff dimensions: the influence of geometry on anastomotic flow patterns
R K Fisher1, T V How, T Carpenter
1Regional Vascular Unit and Department of Clinical Engineering, Royal Liverpool University Hospital, Daulby Street, Liverpool L69 3GA, U.K.
Objectives:
since cuff dimensions are variable, we studied the influence of cuff geometry on flow mechanics, in an attempt to identify the optimum configuration.
Materials And Methods:
bench studies involved the manufacture of anatomically accurate models of varying cuff dimensions, perfused in a specifically designed flow rig, simulating physiological conditions. Flow visualisation studies incorporating laser illumination of tracer particles enabled accurate analysis of flow patterns.
Results:
the vortex created within the proximal cuff of each model during the deceleration phase of the cardiac cycle was strongly influenced by the aspect ratio (AR=cuff height:length). The standard and high cuffs (AR=1.63 and 1.18, respectively) demonstrated cohesive vortices and stable flow patterns. Low and long cuffs (AR=2.6 and 2.25, respectively) created more complex vortices with large areas of flow separation and low velocities.
Conclusions:
aspect ratio has an important influence on flow within the distal anastomosis, with cuff dimensions of 13 mm long and 8-11 mm high (standard and high cuffs) creating beneficial flow patterns anticipated to optimise wall shear stress and inhibit myointimal hyperplasia.