[Evaluation of immunochromatography method for rapid detection of Shigatoxin]
M Kurokawa1, T Miyata, M Murase
1Department of Bacteriology, Kobe Institute of Health.
Abstract:
The Shigatoxin detection kit based on the immunochromatography system is commercially available. To obtain the identification result rapidly, we devised the improved method (ICG-Imp) replaced to an original method (ICG). Modification provided that Shigatoxins extracted directly from the strains grown on TSI medium without centrifugation. ICG-Imp was compared with ICG, RPLA and PCR. Comparing with RPLA, the sensitivity, specificity, and concordance rate of Shigatoxin 1 showed 77.5, 100 and 90.5%, respectively on ICG, 93.8, 100 and 97.4%, respectively on ICG-Imp, and 100, 99.1 and 99.5%, respectively on PCR. On the other hand, the patterns of Shigatoxin 2 showed 95.3, 100 and 96.3%, respectively on ICG, 100, 100, and 100%, respectively on ICG-Imp, and 100, 100 and 100%, respectively on PCR. The time required from TSI medium to the final result are 24 h, 30-60 min, 48 h, and 6 h, respectively by ICG, ICG-Imp, RPLA, and PCR. It seems that the ICG-Imp is recommended for the identification by means of the accuracy and rapidness.