Related Experiment Video
Updated: Aug 9, 2026

Automated Hydrophobic Interaction Chromatography Column Selection for Use in Protein Purification
Published on: September 21, 2011
Ranitidine hydrochloride: development of an isocratic stability--indicating high-performance liquid chromatographic
1Atrix Laboratories, Fort Collins, CO 80525-4417, USA.
Abstract:
The development of a stability-indicating assay for ranitidine hydrochloride using a mobile phase added ion-interaction reagent was achieved. The assay easily separated all known and unknown impurities/degradants. This assay may be used for the determination of purity, identity and strength for the active ingredient and finished dosage forms. Placebo samples were analyzed for all of the dosage forms and did not interfere with the separation. The system was found to be linear over a range of 0.056 to 44.4 microg/g, with a limit of detection of 0.028 microg/g and a limit of quantitation of 0.056 microg/g. The system precision was determined to be 0.7%. The development of the stability-indicating assay and the effect that each chromatographic variable had on the separation will be discussed.
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors
Ion-Exchange Chromatography
Affinity Chromatography