Related Experiment Videos
Potentiometric detection of organic acids in liquid chromatography using polymeric liquid membrane electrodes
D Zielinska1, I Poels, M Pietraszkiewicz
1Warmia and Masuria University in Olsztyn, Department of Chemistry, Poland.
Abstract:
Potentiometric detection employing coated-wire electrodes was applied to the determination of organic acids in liquid chromatography (LC). Poly(vinyl chloride)-based liquid membranes, incorporating lipophilic macrocyclic hexaamines as neutral ionophores were used as electrode coatings. The selectivity and sensitivity of the macrocycle-based electrodes were found to be superior to an electrode based on a lipophilic anion exchanger (a quaternary ammonium salt). Sensitive detection was obtained for the di- and tricarboxylic acids tartaric, malonic, malic, citric, fumaric, succinic, pyruvic, 2-oxoglutaric and maleic acids after separation in reversed-phase LC. Detection limits (signal/4sigmanoise=3) of 6 pmol for malonic acid and 2 pmol for maleic acid were attained. The detection was explained using a molecular recognition model. The hexaamine-based potentiometric electrodes had a 1-s response time at 1 ml min(-1) flow-rates. They were stable for at least 4 months, with an intra-electrode variation of 3.2% (n=5).