Effect of increased repetition time TR on precision of inversion-recovery T(1) measurements
1Department of Radiology, North Shore University Hospital, 300 Community Drive, Manhasset, NY 11030, USA. kingsley@nshs.edu
Abstract:
The effect of increased repetition time, TR, on the precision of inversion-recovery measurements of the spin-lattice relaxation time, T(1), was calculated theoretically, simulated numerically, and measured experimentally. All three methods yielded similar results. With constant inversion times, the T(1) precision was independent of TR. Therefore, 1) multiple-slice inversion-recovery fast-spin-echo T(1) maps should be made one slice at a time, not with interleaved acquisition, and 2) once the longest inversion time t(i) has been set, TR should be set just enough longer than the longest t(i) to allow data acquisition.
Related Concept Videos
Atomic Nuclei: Types of Nuclear Relaxation
In spin–lattice or longitudinal relaxation, the excited spins exchange energy with the surrounding lattice as they return to the lower energy level. Among several mechanisms that contribute to spin–lattice relaxation, magnetic dipolar interactions are significant. Here, the excited nucleus transfers...
NMR Spectrometers: Resolution and Error Correction
Insensitive Nuclei Enhanced by Polarization Transfer (INEPT)
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...
Magnetic Resonance Imaging