Evaluation of in vivo dose measurements for patients undergoing electron boost treatments
1Medical Physics Department, Cookridge Hospital, Hospital Lane, Cookridge, LS 16 6QB, Leeds, UK.
Abstract:
This study evaluated p-type silicon diodes for use in in vivo dosimetry in clinical electron beams. A calibrated p-type silicon diode detector was used to measure the dose received by the patient in the centre of the field. Readings were corrected for energy, temperature and stand-off of the electron applicator from the patient surface. The mean difference between measured and prescribed dose was 1.04% (95% CI 0.72 to 1.36 %).