An evaluation of blood creatinine measurement by creatinase on the NOVA M7 blood gas analyzer
D J Rowe1, H Omar, S L Barratt
1Department of Chemical Pathology, Southampton University Hospitals, NHS Trust, SO16 6YD, Southampton, UK. david.rowe@suht.swest.nhs.uk
Abstract:
This study compared a creatininase method for the analysis at the bedside of creatinine (CR) in whole blood on a NOVA Biomedical M7 analyser, with rate-Jaffe and creatininase-based laboratory methods. Correlation and precision data were obtained and the effect of increased bilirubin concentration was assessed.