Evaluation of CoaguChek Pro ACT in cardiac catheterization laboratory

M Müller-Bardorff1, R H Strasser, S Schön

  • 1Department of Medicine II, University of Lübeck, 23562, Lübeck, Germany.

Insights

The CoaguChek Pro ACT assay effectively monitors heparin in cardiac catheterization labs, showing excellent agreement and good correlation with established methods. Its analytical performance is reliable for activated clotting time (ACT) measurements.

Area of Science:

  • Clinical Chemistry
  • Hematology
  • Medical Devices

Background:

  • Accurate monitoring of anticoagulation, particularly heparin, is critical in cardiac catheterization laboratories (cathlabs).
  • Existing methods for activated clotting time (ACT) measurement require validation for new point-of-care devices.

Purpose of the Study:

  • To evaluate the analytical performance of the CoaguChek Pro ACT assay.
  • To compare CoaguChek Pro ACT assay with the Hemochron Celite ACT assay for activated clotting time (ACT) determination.
  • To assess the correlation of both assays with heparin levels.

Main Methods:

  • Analytical performance assessment of CoaguChek Pro ACT assay versus Hemochron Celite ACT assay.
  • Inclusion of 158 patients and 101 normal subjects across five cathlabs.
  • Comparison of two CoaguChek Pro ACT lots against different Hemochron Celite ACT lots using arterial and venous blood.
  • ACT measurements performed before, directly after, and 1-4 hours post-heparinization.

Main Results:

  • Good correlations were observed between CoaguChek Pro ACT and Hemochron Celite assays for both arterial and venous blood (r=0.88 and 0.84).
  • Excellent agreement was found between the two CoaguChek Pro ACT lots (r=0.99).
  • Correlations with heparin levels were similar for both assays, with minimal influence from hematocrit and platelets; method imprecision was very good (CV<6%).

Conclusions:

  • The CoaguChek Pro ACT assay demonstrates reliable analytical performance for ACT monitoring.
  • The assay is suitable for monitoring heparin in cathlab settings.
  • CoaguChek Pro ACT assay offers a viable alternative for point-of-care ACT testing.