Structural stiffness and reducibility of external fixators placed in malalignment and malrotation

D B Carmack1, K L Kaylor, M J Yaszemski

  • 1Department of Orthopaedic Surgery, Wilford Hall USAF Medical Center, Lackland Air Force Base, Texas 78236-53000, USA.

Summary

This study found that not all external fixators can achieve anatomic reduction in malaligned fractures without pin repositioning. External fixator design significantly impacts structural stiffness, crucial for optimal fracture immobilization.