Convenient dioxin measuring method using an efficient sampling train, an efficient HPLC system and a highly sensitive HRGC/LRMS with a PTV injector
Related Experiment Video
Updated: Oct 7, 2026

Tuning a Parallel Segmented Flow Column and Enabling Multiplexed Detection
Published on: December 15, 2015
K Urano1, M Kato, Y Nagayanagi
1Laboratory of Safety and Environmental Engineering, Yokohama National University, Japan. k-urano@ynu.ac.jp
Abstract:
An efficient sampling method for dioxins from flue gas, an efficient automatic purification and fractionation method by a new HPLC system using a sulfuric acid-silica/silica column, a Nitro column and a PGC column, and sensitive determination method by an economical HRGC/LRMS using a programmable temperature vaporization (PTV) injector were developed. It was confirmed that the dioxins could be easily collected by the proposed simple sampling train consisting of only two bottles, and the extracted sample could be purified sufficiently and separated as mono-ortho PCBs, non-ortho PCBs and PCDDs/PCDFs completely with the proposed HPLC system. The peak areas of GC/MS were increased linearly with the injection volume up to 40 microl by the PTV technique, and the sensitivity could be increased to 20 times higher than usual. This convenient measuring method can drastically reduce operation time, consumption of hazardous solvent and cost.
More Related Videos
Related Concept Videos
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Gas Chromatography: Types of Detectors-II