Electrocautery-induced tachycardia in a rate-responsive pacemaker.
1Department of Anesthesiology, University of Toronto, Canada. david.wong@uhn.on.ca
Electrocautery can interfere with on-demand pacemakers, causing them to pace unexpectedly. In a VVIR mode pacemaker, this interference resulted in pacing at the maximum programmed rate of 130 pulses/min.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- On-demand pacemakers are susceptible to electromagnetic interference from electrocautery.
- Such interference can lead to pacemaker inhibition or inappropriate pacing, posing risks to patients.
Observation:
- A patient with a pacemaker experienced electrical interference during a medical procedure.
- The interference was specifically caused by electrocautery use.
- The pacemaker was operating in ventricle-paced, ventricle-sensed, inhibited, rate-responsive (VVIR) mode.
Findings:
- Electrocautery induced inhibition and loss of pacing in the affected pacemaker.
- The VVIR pacemaker was triggered to pace at its programmed maximum rate of 130 pulses/min.
- This demonstrates a specific failure mode of pacemakers under electrocautery interference.
Implications:
- Highlights the risk of electrocautery interference with VVIR pacemakers.
- Suggests the need for careful monitoring and potential adjustments when using electrocautery in patients with pacemakers.
- Informs strategies for mitigating pacemaker malfunction during surgical procedures involving electrocautery.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
10:08Tachycardia-Induced Cardiomyopathy As a Chronic Heart Failure Model in Swine
Published on: February 17, 2018
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias III: Characteristics of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Mechanism of Cardiac Arrhythmias
Electrophysiology of Normal Cardiac Rhythm