Characterization of aerosol output from various nebulizer/compressor combinations
C Reisner1, R K Katial, B B Bartelson
1Allergy and Immunology, National Jewish Center for Immunology and Respiratory Medicine, Denver, CO, USA. cr33898@glaxowellcome.com
Objectives:
Different commercially available nebulizers and compressors are available. However, the optimal combination for drug delivery is unknown.
Methods:
Flow rates of five different compressors (n = 3/compressor) tested alone and in combination with five different commercial nebulizers (n = 9 of each brand of nebulizer) were evaluated. Thereafter, the performances of the different nebulizers were evaluated using 2.5 mg albuterol solution (0.5 mL) added to 2.5 mL saline at flow rates of 2, 3, 4, and 5 L/minute using a laser particle analyzer. Volume median diameter and percentage of particles in the respirable range (1-5 microm) were calculated from this data. Time for nebulization (in seconds) and residual volume (in milliliters) were also recorded.
Results:
The mean flow rates for the compressors evaluated without a nebulizer attached ranged from 6.6 L/minute (LifeCare Freedom-neb; LifeCare International, Lafayette, CO) to 12.2 L/minute (DeVilbiss Pulmo-Aide; DeVilbiss Health Care, Somerset, PA). Flow rates for the nebulizer/compressor combinations ranged from 2.08 L/minute (Pari LC Jet Proneb; Pari Respiratory Equipment, Richmond, VA) to 5.42 L/minute (Puritan Bennett Raindrop; Puritan Bennett, Lenexa, KS/Omron Compare; Omron, Health Care,Vernon Hills, IL). Using the repeated measure ANOVA model, the interaction between flow rate and device was significant (P < 0.001) for both percentage of particles in the respirable range and log volume median diameter. It was observed that the percentage of particles in the respirable range for the Pari LC Jet did not increase across flow rates in contrast to the other 4 nebulizers. All comparisons to the Pari LC Jet at 2 L/minute were significant.
Conclusions:
Marked variability exists in the flow rates among different commercially available compressors used for home nebulization of inhaled pulmonary medications. Different nebulizer/compressor combinations have markedly different performance characteristics which could result in different efficacy and safety profiles of the medications being administered via these devices. We recommend that this type of information be used as a starting point for selecting different nebulizer/compressor combinations. Further clinical evaluation is warranted.
More Related Videos
Related Concept Videos
Inhaled Medications
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...