Recombinant Microdochium nivale carbohydrate oxidase and its application in an amperometric glucose sensor
J Kulys1, L Tetianec, P Schneider
1Institute of Biochemistry, Mokslininku 12, 2600 Vilnius, Lithuania. jkulys@bchi.lt
Abstract:
Biosensors containing recombinant carbohydrate oxidase from Microdochium nivale (rMnO) were developed by means of either chemically modified carbon paste or graphite electrode. 1-(N,N-dimethylamine)-4-(4-morpholine)benzene (AMB) and 1,1'-dimethylferrocene (DMFc) have been used as mediators. The biosensors showed a linear calibration graph up to 18 mM of glucose when operated at 0.04-0.36 V versus a saturated calomel electrode. Almost no change was detected in the sensitivity of the biosensors at pH 7.2-8.1. The biosensors responded to other aldoses in the D-configuration, however, maximal sensitivity of the biosensor was towards D-glucose. The biosensor did not response to polyhydroxylic compounds such as D-mannitol, D-sorbitol and inositol. The advantages of the biosensors based on rMnO in comparison to Aspergillus niger glucose oxidase is a wider linear range, low sensitivity to oxygen and (in some cases) broad specificity.