Correction of helical CT attenuation values with wide beam collimation: in vitro test with urinary calculi
J C Williams1, K C Saw, A G Monga
1Department of Anatomy and Cell Biology, Indiana University School of Medicine, Indianapolis 46202, USA.
Rationale And Objectives:
Urinary calculi are now commonly detected with helical computed tomography (CT), and it has been proposed that stone composition can be determined from CT attenuation values. However, typical scans are made with a beam collimation of 5 mm or more, resulting in volume averaging and reduction in accuracy of attenuation measurement. The authors tested a model for correction of errors in attenuation values, even at section widths larger than the width of the object.
Materials And Methods:
Human urinary stones were scanned with helical CT at different beam collimation widths. A computer model was used to predict the effect of beam width and stone size on accuracy of measured attenuation.
Results:
At 3-mm collimation, the model corrected the attenuation readings with an underestimation of 12% +/- 1 (compared with values at 1-mm collimation; 127 stones; diameters of 1.7-11.3 mm). With attenuation measured at 10-mm collimation, the model underestimated the true value by 34% +/- 3 (103 stones), with a significant negative correlation with stone diameter on magnitude of error (diameters of 3.0-11.3 mm). Correlation of data from patient scans with subsequent in vitro scanning of the same stones confirmed the validity of the model, but corrected in vivo scans consistently yielded lower values for the stones than in vitro.
Conclusion:
Volume averaging effects on attenuation in helical CT are predictable in vitro for urinary calculi--and presumably for other roughly spherical structures--as long as section width does not excessively exceed the diameter of the structure.
More Related Videos
07:12Semiautomated Longitudinal Microcomputed Tomography-based Quantitative Structural Analysis of a Nude Rat Osteoporosis-related Vertebral Fracture Model
Published on: September 28, 2017
12:39DUCT: Double Resin Casting followed by Micro-Computed Tomography for 3D Liver Analysis
Published on: September 28, 2021
Related Concept Videos
Urinary Tract Calculi I: Introduction
Urinary Tract Calculi III: Medical Management
Urinary Tract Calculi V: Nursing Management
Imaging Studies II: Ultrasonography
Imaging Studies III: Computed Tomography
Imaging Studies V: Intravenous Urography and Retrograde Pyelography