Evaluation of the FASTSTART mode for reducing start-up delay in syringe pump infusion systems
Objective:
The aim of the study was to evaluate the IVAC P7000 FASTSTART mode with regard to start-up performance in a 50-ml infusion syringe at a flow rate of 1 ml.h-1.
Methods:
The time from depression of the start button to first fluid flow (T1) and to establishment of a pre-set flow rate (T2) were gravimetrically recorded with and without FASTSTART and with and without priming of the infusion system with a 1-ml fluid bolus prior to connection of the infusion line to the patient.
Results:
FASTSTART significantly reduced start-up times in the unprimed syringe pump infusion system from (mean [SD]) 9.4 (6.0) to 2.5 (3.5) min for T1 and from 21.8 (9.8) to 9.4 (6.2) min for T2 (all p < 0.001). The greatest improvement in shortening of T1 and T2 was obtained when the system was primed prior to starting (p < 0.0001). After priming the infusion system, FASTSTART shortened T2 by some 50% from 1.4 (1.4) to 0.7 (0.6) min.
Conclusion:
Our data indicate that the FASTSTART procedure is effective and that substantial improvements can be obtained by priming the system prior to starting.