Sub-miniature ExB sector-field mass spectrometer.

J A Diaz1, C F Giese, W R Gentry

  • 1Universidad de Costa Rica, San Jose. jdiaz@micit.go.cr

Journal of the American Society for Mass Spectrometry
|June 13, 2001
PubMed
Summary

Researchers developed a novel miniature mass spectrometer using advanced fabrication techniques. This compact, low-cost instrument offers performance comparable to commercial analyzers, enabling portable analytical applications.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

The dynamics of arsenic and copper in solid and aqueous phases in reactive confluences receiving acid drainage: The role of turbidity and particle size.

Environmental pollution (Barking, Essex : 1987)·2023
Same author

Use of healthcare REsources and associated COsts in controlled versus uncontrolled carcinoid SYndrome in patients with neuroendocrine tumours: the RECOSY study.

Clinical & translational oncology : official publication of the Federation of Spanish Oncology Societies and of the National Cancer Institute of Mexico·2021
Same author

Aerial strategies advance volcanic gas measurements at inaccessible, strongly degassing volcanoes.

Science advances·2020
Same author

First Report of Tomato Yellow Leaf Curl Virus-Is in Spain: Coexistence of Two Different Geminiviruses in the Same Epidemic Outbreak.

Plant disease·2019
Same author

Spanish Melon necrotic spot virus Isolate Overcomes the Resistance Conferred by the Recessive nsv Gene of Melon.

Plant disease·2019
Same author

P-Selectin Inhibition Therapeutically Promotes Thrombus Resolution and Prevents Vein Wall Fibrosis Better than Enoxaparin and an Inhibitor to von Willebrand Factor.

Journal of vascular surgery. Venous and lymphatic disorders·2016

Area of Science:

  • Analytical Chemistry
  • Instrument Development
  • Sensor Technology

Background:

  • Traditional mass spectrometers are often large, expensive, and power-intensive.
  • There is a need for portable, cost-effective analytical instruments for various applications.

Purpose of the Study:

  • To design and fabricate a novel sub-miniature, double-focusing sector-field mass spectrometer.
  • To achieve high performance in a compact and low-power device.
  • To enable portable and low-cost analytical instrumentation.

Main Methods:

  • Fabrication using conventional machining and thin-film patterning.
  • Design based on a 90-degree cylindrical crossed electric and magnetic sector-field analyzer.
  • Performance optimization using finite element analysis and computer modeling.

Related Experiment Videos

Main Results:

  • Achieved a resolving power of 10^6 (FWHM).
  • Demonstrated a detection limit near 10 parts per million and a dynamic range of 5 orders of magnitude.
  • Mass range up to 10^3 Da with a compact size (3.5x6x7.5 cm), low weight (0.8 kg), and low power consumption (2.5 W).

Conclusions:

  • The miniature mass spectrometer exhibits performance comparable to commercial residual gas analyzers.
  • The instrument's small size, low cost, and high performance make it suitable for portable analytical applications.
  • This development opens possibilities for widespread use in field analysis and resource-limited settings.