Related Experiment Video
Updated: Aug 12, 2026

Identification of Olfactory Volatiles using Gas Chromatography-Multi-unit Recordings (GCMR) in the Insect Antennal Lobe
Published on: February 24, 2013
Miniature radio-frequency mobility analyzer as a gas chromatographic detector for oxygen-containing volatile organic
G A Eiceman1, B Tadjikov, E Krylov
1Department of Chemistry and Biochemistry, New Mexico State University, Las Cruces 88003-0001, USA. geiceman@nmsu.edu
Abstract:
A high electric field, radio-frequency ion mobility spectrometry (RF-IMS) analyzer was used as a small detector in gas chromatographic separations of mixtures of volatile organic compounds including alcohols, aldehydes, esters, ethers, pheromones, and other chemical attractants for insects. The detector was equipped with a 2 mCi 63Ni ion source and the drift region for ion characterization was 5 mm wide, 15 mm long and 0.5 mm high. The rate of scanning for the compensation voltages was 60 V s(-1) and permitted four to six scans to be obtained across a capillary chromatographic elution profile for each component. The RF-IMS scans were characteristic of a compound and provided a second dimension of chemical identity to chromatographic retention adding specificity in instances of co-elution. Limits of detection were 1.6-55 x 10(-11) g with an average detection limit for all chemicals of 9.4 x 10(-11) g. Response to mass was linear from 2-50 x 10(-10) g with an average sensitivity of 4 pA ng(-1). Separations of pheromones and chemical attractants for insects illustrated the distinct patterns obtained from gas chromatography with RF-IMS scans in real time and suggest an analytical utility of the RF-IMS as a small, advanced detector for on-site gas chromatographs.
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Types of Detectors