[Quality of ambulatory thoracic radiography in the child--a pilot study]
D Engelmann1, T Dütting, R Wunsch
1Abteilung für Pädiatrische Radiologie, Universität Heidelberg.
Abstract:
A quality control of outpatient paediatric chest X-rays was conducted in a sample of patients of one paediatric practice. During a period of eight months the technical image quality was analysed considering both diagnostic aspects and radiation protection. The quality of the 139 examined chest X-rays was inadequate concerning the collimation and focussing of the X-rays and the positioning of the patients. Exposure was estimated as average, sharpness was rated as good. In total 14% of the X-rays were not suitable for medical diagnosis. Image quality of the X-rays of infants (children younger than 6 years) was significantly lower compared to the total sample. Radiation protection standards were not fulfilled. As a conclusion from our results, improvements in outpatient paediatric radiography are urgently necessary. Quality control committees should pay particular attention in radiographs of infants.