Prolonged interference of patent blue on pulse oximetry readings
1Department of Anesthesiology, Kaohsiung Veterans General Hospital, 386, Ta-Chung 1st Road, Kaohsiung 813, Taiwan, R.O.C. yychia@isca.vghks.gov.tw
Background:
We evaluated the effect of intra-arterial injection of Patent Blue (PB) on pulse oximetry.
Methods:
Ten consecutive female patients who underwent intra-arterial insertion of catheter for chemotherapy were enrolled in this study. All patients received general anesthesia and ECG, pulse oximetry, and direct arterial blood pressure were monitored throughout the perioperative period. Baseline arterial blood saturation (SpO2), haemoglobin concentration (Hgb), estimated arterial saturation (SaO2), arterial oxygen partial pressure (PaO2) were obtained through arterial blood gas analysis, and true arterial saturation (ScoO2) through CO-Oximetry analysis. The subsequent values obtained at various points of time following the intra-arterial administration of PB were compared with the baseline data.
Results:
The SpO2 readings taken after intra-arterial injection of PB were significantly lower than the baseline reading. The extent of the reduction of SpO2 readings two hours after the administration of PB was relevant to the total accumulative dosage of PB (rs = -0.846, P < 0.05). However, the reduction of SpO2 bore no significant relation with Hgb concentration or with the ratio of PB dosage to Hgb concentration.
Conclusions:
Intra-arterial administration of PB interferes with pulse oximetry readings and the effect lasts for a variable duration. The extent of desaturation may be related to the total accumulative dosage of intra-arterial PB injected.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
04:44Near-Infrared Spectroscopy During Reactive Hyperemia for the Assessment of Lower Limb Vascular Function
Published on: March 22, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...