Assay of ochratoxin A in wine and beer by high-pressure liquid chromatography photodiode array and gas chromatography
G J Soleas1, J Yan, D M Goldberg
1Quality Assurance Department, Liquor Control Board of Ontario, 55 Lakeshore Boulevard East, Toronto, Ontario, Canada M5E 1A4.
Abstract:
To routinely assay the concentrations of ochratoxin A (OTA) in wines and beers, two new methods were developed and evaluated. The first utilized solid-phase extraction on a C(18) cartridge to achieve a 100-fold sample concentration followed by high-performance liquid chromatography on a C(18) column with gradient elution and quantitation at 333 nm by means of a photodiode array detector. Positive confirmation can be carried out by purity and match-factor analysis as well as peak shift following esterification with BF(3). Total run time is 28 min. The limits of detection (LOD) and quantitation (LOQ) are 0.05 and 0.10 microg/L, respectively. Recovery and imprecision ranged from 83 to 94% and from 4.0 to 8.9%, respectively. With a throughput of 35 assays per working day, this method is ideal for routine OTA analysis. It was used to survey the concentrations of OTA in 942 wines (2 of which gave values between 0.1 and 0.2 microg/L) and 107 beers (2 of which gave values between 0.05 and 0.1 microg/L). OTA was detected more frequently in red than white wines, with the highest incidence in red wines from Spain and Argentina. There was no association between OTA and country of origin or beverage type among the beers analyzed. The second method utilized gas chromatography with mass selective detection monitoring eight specific ions, preceded by extraction in dichloromethane and derivatization with bis[trimethylsilyl]trifluoroacetamide. LOD and LOQ were 0.1 and 2 microg/L, respectively; recovery and imprecision were 69-75 and 9.0-11.1%, respectively. The method is not suitable for routine quantitation but is potentially useful as a confirmatory tool for samples with OTA > or =0.1 microg/L.
More Related Videos
08:43PTR-ToF-MS Coupled with an Automated Sampling System and Tailored Data Analysis for Food Studies: Bioprocess Monitoring, Screening and Nose-space Analysis
Published on: May 11, 2017
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
High-Performance Liquid Chromatography: Types of Detectors