Related Experiment Videos
Non-uniformity of contact stress on polyethylene inserts in total knee arthroplasty
RE Bristol1, DC Fitzpatrick, TD Brown
1Departments of Orthopaedic Surgery, Mechanical Engineering, and Biomedical Engineering, University of Iowa, Iowa City IA, USA.
Abstract:
This study examined tibiofemoral contact stresses in 15 commercially available TKR designs, using digitally imaged pressure-sensitive film. The objectives were (1) to determine the correlation between minimization of spatial mean contact stress and minimization of the amount of overloaded (>10 MPa) polyethylene contact area, and (2) to ascertain the difference in contact stresses for machined versus moulded polyethylene inserts. The data showed that there was no statistically significant (design rank-order) correlation of spatial mean contact stress with overloaded polyethylene area. The data also showed that machining cutter preparation of the polyethylene insert caused substantial local contact stress non-uniformities that corresponded to the pattern of grossly visible machining marks.