The carbon dioxide absorption capacity of Amsorb is half that of soda lime
H Higuchi1, Y Adachi, S Arimura
1Department of Anesthesia, Self Defense Force Central Hospital, Tokyo, Japan. higu-chi@ka2.so-net.ne.jp
Unlabelled:
A new CO(2) absorbent, Amsorb (A), which does not contain monovalent bases, is ideal because it does not degrade volatile anesthetics to either Compound A (from sevoflurane) or carbon monoxide (from desflurane, enflurane, or isoflurane). The CO(2) absorption capacity of A, however, has not been investigated under clinical conditions. In this study, we compared the longevity (time to exhaustion) and CO(2) absorption capacity (the volume of CO(2) absorbed before CO(2) rebreathing occurs) of A under low-flow anesthesia (1 L/min) with those of two soda lime absorbents-Medisorb (M) and Sodasorb (S)-by using a 750-mL ADU canister and a 1350-mL Aestiva 3000 canister. In the study with the ADU canister, the longevity of A was 213 +/- 71 min, significantly less than those of M (445 +/- 125; P < 0.01) and S (503 +/- 89; P < 0.001). The CO(2) absorption capacity (L/100 g absorbent) of A was 5.5 +/- 1.2, significantly less than those of M (10.7 +/- 1.7) and S (12.1 +/- 1.8; P < 0.001). In the study with the Aestiva 3000 canister, the longevity of A was 218 +/- 61 min, significantly less than those of M (538 +/- 136) and S (528 +/- 103; P < 0.001). The CO(2) absorption capacity (L/100 g absorbent) of A was 7.6 +/- 1.6, significantly less than those of M (14.4 +/- 1.8) and S (14.8 +/- 2.3; P < 0.001). These results indicate that the CO(2) absorption capacity of A is half that of M or S and that the difference in the CO(2) absorption capacity between A and M or S is almost constant, regardless of the canister design.
Implications:
The CO(2) absorption capacity of Amsorb is half that of Medisorb and Sodasorb under clinical low-flow (1 L/min) anesthesia with either a 750-mL Ohmeda ADU compact or a 1350-mL Ohmeda Aestiva 3000 canister.
More Related Videos
11:14Two-way Valorization of Blast Furnace Slag: Synthesis of Precipitated Calcium Carbonate and Zeolitic Heavy Metal Adsorbent
Published on: February 21, 2017
09:46Adsorption Device Based on a Langatate Crystal Microbalance for High Temperature High Pressure Gas Adsorption in Zeolite H-ZSM-5
Published on: August 25, 2016
Related Concept Videos
Carbonation Shrinkage
The concrete's permeability is slightly reduced as calcium carbonate produced during the reaction fills its pores. Furthermore, its strength is slightly enhanced as the water released during the reaction facilitates the...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Loss of Carboxy Group as CO2: Decarboxylation of Malonic Acid Derivatives
The Calvin Benson Cycle
Bicarbonate-Carbonic Acid Buffer
Hydration of Cement