Collection efficiency and capacity of three samplers for acidic and basic gases
Related Experiment Video
Updated: Oct 5, 2026

Calibrated Passive Sampling - Multi-plot Field Measurements of NH3 Emissions with a Combination of Dynamic Tube Method and Passive Samplers
Published on: March 21, 2016
C J Tsai1, C H Huang, S H Wang
1Institute of Environmental Engineering, National Chiao Tung University, Hsin Chu, Taiwan. cjtsai@cc.nctu.edu.tw
Abstract:
This paper compares the collection efficiency and capacity of a coated porous metal disk with those of a silica gel tube and an impinger at the acid or basic aerosol concentration of two times the permissible exposure limitation (Taiwan IOSH, Institute of Occupational Safety and Health), relative humidity of 80 +/- 5%, and temperature of 30 +/- 3 degrees C in the laboratory. Using 5% sodium carbonate/glycerin coating, the collection efficiencies of the porous metal disk for nitric acid, hydrogen chloride, and hydrogen fluoride are found to be higher than those of the silica gel tube and the impinger when the sampling time is less than 3.0 h. The gas collection capacity of the porous metal disk coated with 5% sodium carbonate is calculated to be 3.9, 5.8, and >2.5 mg for nitric acid, hydrogen chloride, and hydrogen fluoride, respectively. For ammonia, the porous metal disk coated using 4% citric acid performs slightly better than the impinger, and the corresponding capacity is slightly greater than 33.6 mg.
Related Concept Videos
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...
Gas Chromatography: Types of Detectors-II
Atomic Emission Spectroscopy: Instrumentation
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...