Accuracy of aneroid sphygmomanometers in clinical usage: University of Michigan experience
1Division of Hypertension, University of Michigan Health System, Chelsea, Michigan 48118, USA. syarows@umich.edu
Background:
Aneroid manometers are frequently used to measure blood pressure. Aneroid manometers have moving parts that are subject to fatigue. The accuracy duration of the aneroid devices, like most digital devices, is unstudied. It has been accepted that if the aneroid device does not rest at '0' it is inaccurate, but how often is the device inaccurate when it does rest at '0'?
Methods:
A Universal Biometer DPM-III measuring unit was used for all of the measurements at 10 University of Michigan Health System sites. A total of 136 aneroid manometers were tested. Two additional aneroid devices were not tested, as the needle did not start within '0'. Static pressure measurements were made at nine levels for all devices: 50, 80, 90, 100, 120, 150, 200 and 250mmHg.
Results:
The average difference of the nine pressure settings of the whole group was 0.2+/-0.31 (95% confidence interval 0.1-0.2) mmHg. The largest number of devices that were not calibrated within +/-3mmHg was seen at the 150mmHg setting with six (4.4%) of the devices failing. If an accuracy standard of +/-2mmHg was used, the largest number of devices failed at 250mmHg (22 devices, 16.2%). The largest number of devices that were inaccurate was manufactured 6 years prior to testing and was from two sites.
Conclusions:
Aneroid devices were accurate. A yearly calibration programme should be performed and a +/-2mmHg standard should be used. Portable aneroid manometers may need to be more frequently calibrated due to the trauma associated with dropping.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.