Related Experiment Video
Updated: Oct 5, 2026

Simple In-House Ultra-High Performance Capillary Column Manufacturing with the FlashPack Approach
Published on: December 4, 2021
Capillary-electrochromatographic separations with copolymeric reversed-stationary phase and ion-exchanger-packed
1University of Iowa, Department of Chemistry, Iowa City 52242, USA.
Abstract:
A macroporous, spherical, 7 microm, polystyrene-divinylbenzene (PS-DVB), reversed-phase adsorbent (PRP-1) was evaluated as a stationary phase for the capillary electrochromatographic (CEC) separation of neutral, acidic, and basic analytes of pharmaceutical interest. Electroosmotic flow (EOF) for a PRP-1 packed capillary is nearly constant over the pH 2 to 10 range and is higher than for a silica-based C18 packed capillary on the acidic side. EOF increases with an increase in buffer acetonitrile concentration or as applied potential increases. As analyte hydrophobicity increases, analyte retention and migration time increases. Increasing buffer acetonitrile concentration reduces analyte partitioning with the PS-DVB stationary phase and analyte retention and migration time decreases. When exchange sites are present on the PS-DVB copolymer, EOF (EOF is reversed for the anion-exchanger) increases as the exchange capacity increases. An increased exchange capacity also reduces partitioning of the analyte with the PS-DVB matrix and analyte retention and migration time decrease. Because of excellent stability in an acid environment, the PRP-1 packed capillary can be used in strong acid buffer solution and weak acid and base analytes depending on pKa values can be separated as neutral species and cations, respectively. CEC separations on a PRP-1 capillary of neutral steroids, weak base pharmaceuticals (separation as cations), purines and pyrimidines (as cations), fatty acids (as undissociated species), and sulfa derivatives (as cations) are described. Efficiency for the PRP-1 packed capillary for acetone or thiourea as the analyte is about 6 x 10(4) plates m(-1).
Related Concept Videos
Ion-Exchange Chromatography
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Electrophoresis: Overview
There...
Ion Exchange
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Capillary Electrophoresis: Instrumentation