Validation of ethylene oxide-sterilization using a new measuring system
Related Experiment Video
Updated: Oct 5, 2026

Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
Abstract:
Up to now the parametric release of medical devices sterilized with EO was not possible due the lack of an online method of measurement for EO. With the developed mobile system it is possible to measure all physical parameters of the sterilization process online in the chamber. The accuracy of the EO concentration and humidity measurement is < +/- 1%, of the temperature < +/- 0.05 degree C and of pressure < 0.5% of the measured value.