Dipstick measurements of urine specific gravity are unreliable
A S de Buys Roessingh1, A Drukker, J P Guignard
1Renal Unit, Department of Paediatrics, University Medical Centre (CHUV), CH-1011 Lausanne, Switzerland.
Aim:
To evaluate the reliability of dipstick measurements of urine specific gravity (U-SG).
Methods:
Fresh urine specimens were tested for urine pH and osmolality (U-pH, U-Osm) by a pH meter and an osmometer, and for U-SG by three different methods (refractometry, automatic readout of a dipstick (Clinitek-50), and (visual) change of colour of the dipstick).
Results:
The correlations between the visual U-SG dipstick measurements and U-SG determined by a refractometer and the comparison of Clinitek((R))-50 dipstick U-SG measurements with U-Osm were less than optimal, showing very wide scatter of values. Only the U-SG refractometer values and U-Osm had a good linear correlation. The tested dipstick was unreliable for the bedside determination of U-SG, even after correction for U-pH, as recommended by the manufacturer.
Conclusions:
Among the bedside determinations, only refractometry gives reliable U-SG results. Dipstick U-SG measurements should be abandoned.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Uncertainty in Measurement: Reading Instruments
Urine: Physical and Chemical Properties
The concentration of these solutes varies, with urea being the most abundant nitrogenous waste product. Other solutes include sodium, chloride, potassium, phosphate,...
Specific Gravity of Aggregate
Bulk or gross specific gravity is calculated by taking the ratio of the mass of aggregates in the saturated surface-dry state to the total volume that includes both the solids and the voids within the...
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Urine Studies I: Urinalysis