An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Mechanical testing of bioresorbable implants for use in metacarpal fracture fixation
K J Bozic1, L E Perez, D R Wilson
1Orthopaedic Biomechanics Laboratory, Beth Israel Deaconess Medical Center, Boston, MA, USA.
Abstract:
The purpose of this study was to evaluate the mechanical properties of a bioresorbable plate and screw system that was developed for the treatment of unstable metacarpal fractures and to compare the strength and stiffness of this system with those of conventional titanium plates and screws. Using a 4-point bending test, we measured the strength and stiffness of these implants over a 12-week period of in vitro degradation. Our data suggest that these implants provide stable bending strength and stiffness for 8 weeks and gradually lose their strength over a period of 12 weeks. Further research is necessary to determine whether this level of fixation is adequate to stabilize unstable metacarpal fractures.