Related Experiment Video
Updated: Oct 5, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Initial clinical experience with a new arrhythmia detection algorithm in dual chamber implantable cardioverter
D Theuns1, A P Klootwijk, G P Kimman
1Department of Cardiology, Thoraxcentre, Rotterdam, The Netherlands.
Aim:
Inappropriate therapy, due to poor discrimination of supraventricular tachycardia (SVT) from ventricular tachycardia (VT) remains a major problem in patients with an implantable cardioverter defibrillator (ICD). Theoretically, the addition of atrial sensing in discrimination algorithms should improve this differentiation. The aim of the study is to evaluate the performance of a new tachycardia discrimination algorithm, SMART Detection.
Methods And Results:
Twenty-six patients received a non-thoracotomy ICD system (Phylax AV, Biotronik, Germany). All documented spontaneous arrhythmia episodes were analyzed. During a mean follow-up of 8 months, a total number of 139 events with stored electrograms were recorded in 12 patients. The final diagnosis was ventricular fibrillation (VF) or polymorphic VT (n=20), monomorphic VT (n=69), SVT (n=26), other ventricular arrhythmia (n=3) and T wave oversensing (n=21). In 6 episodes a dual tachycardia was present. Considering SVT episodes, inappropriate therapy occurred in 2 cases of atrial flutter due to stable ventricular rate (<30 ms), 1 case of atrial tachycardia and 2 cases of sinus tachycardia due to a sudden onset (> 10%).
Conclusion:
With the SMART Detection algorithm, discrimination of VT from SVT achieved a sensitivity of 100%, with an accuracy of 95.6% for all ventricular arrhythmias. In the case of SVT, the algorithm appropriately detected and inhibited therapy in 88% of atrial fibrillation.
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiomyopathy II: Dilated Cardiomyopathy
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiopulmonary Resuscitation III: AED Use
Dysrhythmias I: Introduction
Dysrhythmias VI: Management of Dysrhythmias