Unexpected ICD pulse generator failure due to electronic circuit damage caused by electrical overstress
R G Hauser1, D L Hayes, A K Almquist
1Minneapolis Heart Institute Foundation, 920 E. 28th Street, Suite 300, Minneapolis, MN 55407, USA. rhauser747@aol.com
Abstract:
Because it is a lifesaving device, the unexpected failure of an ICD can be catastrophic. We report ICD electronic circuit failure due to electrical overstress damage (EOS) to the high voltage hybird circuit and other electronic components in a series of ICD pulse generator models. Data were obtained from the Multicenter Registry of Pacemaker and ICD Pacemaker and Lead Failures, and from the manufactures' adverse event reports, that were in the FDA's Manufacturer and User Facility Device Experience (MAUDE) database. Of 16 nonbattery Guidant/CPI ICD pulse generator failures reported to the registry, 6 (38%) have been confirmed by the manufacturer to be EOS related, and Guidant/CPI has reported 273 such failures to the FDA as of 12/29/00. The signs of failure included loss of telemetry and inability to deliver therapy, and some patients have experienced serious adverse events. Hybrid circuit damage may have occurred during capacitor charging or reform, and the majority appears to have happened during normal ICD function. While the incidence of this problem is unknown, a management strategy should be adopted that includes routine follow-up every 3 months and device evaluation after a shock or exposure to external defibrillation or electrosurgical devices. This study suggests that additional data are needed to determine the incidence of this problem, and that our present methods for monitoring the performance of ICD's following market release are inadequate.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electric Generator: Alternator
The magnetic flux passing through the coil varies sinusoidally as the loop rotates inside the magnetic field. This...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Power System Three-Phase Short Circuits
Overcurrent Relays
Instantaneous overcurrent relays activate immediately when the input current exceeds a predetermined value, known as the pickup current, instantly energizing the circuit breaker trip coil. This rapid response is vital for addressing severe faults quickly.
Time-delay overcurrent relays, on the other...