Attenuation smear: a 'paradoxical' increase in counts due to attenuation artifact
1Section of Cardiology, Temple University Hospital, Philadelphia, Pennsylvania 19140, USA. CL.Hansen@unix.temple.edu
Background:
Attenuation is a well recognized cause of reconstruction artifacts in SPECT imaging. Occasionally, we have noted an increase in activity extending from the apical septal portion of the ventricle in women with significant breast attenuation. Although the idea that attenuation can produce an increase in activity on the reconstructed images seems paradoxical at first, it is consistent with the process of filtered back projection.
Methods:
We filled a cardiac phantom with 1 mCi of Technetium-99m, placed it in a water filled anthropomorphic torso phantom and imaged it over a 180 degree orbit. Next, a breast phantom designed to simulate a significant degree of breast attenuation was placed on the torso phantom and imaging was repeated. The images were reconstructed first using conventional filtered back projection then with maximum likelihood.
Results:
When the phantoms with and without breast attenuation were reconstructed using filtered back projection and compared, the phantom with breast attenuation had a large 'smear' of activity extending anteriorly from the apical septal wall which was very similar to the abnormalities previously noted in clinical images; the phantom without breast attenuation had no such defect. This artifact was significantly less prominent when the images were reconstructed using the maximum likelihood technique.
Conclusions:
Attenuation artifact can also produce a seemingly paradoxical increase in counts on the reconstructed image but this phenomenon is consistent with the workings of filtered back projection.
Related Concept Videos
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Aliasing
If the sampling frequency is below the Nyquist rate, these replicas overlap, preventing the original signal...