Related Experiment Videos
Sensitive determination of pethidine in body fluids by surface ionization organic mass spectrometry
A Ishii1, R Kurihara, K Watanabe-Suzuki
1Department of Legal Medicine and Bioethics, Nagoya University Graduate School of Medicine, Japan. akishii@med.nagoya-u.ac.jp
Abstract:
We have presented a simple and sensitive method for determining pethidine, a narcotic analgesic drug in body fluids by gas chromatography (GC)/surface ionization organic mass spectrometry (SIOMS). Good linearity was obtained in the range of 0.625-25 ng/ml of whole blood and urine by mass chromatography, and in the range of 0.05-2 ng/ml of whole blood by selected ion monitoring (SIM). Pethidine and diphenylpyraline (internal standard) were extracted from body fluids with Bond Elut Certify cartridges; their recoveries were above 95%. The detection limits (signal-to-noise ratio=3) were estimated to be 0.2 ng/ml of whole blood or urine by mass chromatography, 0.02 ng/ml of whole blood by SIM.