Context:
Unanticipated pacemaker and implantable cardioverter-defibrillator (ICD) generator malfunctions sometimes warrant recall by the US Food and Drug Administration (FDA). Despite increasingly frequent device implantation, pacemaker and ICD recalls and safety alerts (advisories) remain poorly characterized.
Objectives:
To determine pacemaker and ICD generator advisory rates in the United States, to identify trends in these rates, and to examine their clinical and financial implications.
Design And Setting:
Analysis of weekly FDA Enforcement Reports issued between January 1990 and December 2000 to identify all advisories involving pacemaker or ICD generators in the United States. Recalls and safety alerts involving lead malfunctions were not included.
Main Outcome Measures:
Number of pacemakers and ICD generators in the United States subject to FDA recall or safety alert in 1990-2000; annual pacemaker and ICD advisory rates in the United States in 1990-2000; and estimated cost of device advisories.
Results:
During the study period, 52 advisories (median [25th and 75th percentiles], 4 [4 and 7] per year) involving 408 500 pacemakers and 114 645 ICDs (523 145 total devices) were issued. Hardware malfunctions (35 advisories affecting 280 641 devices) and computer errors (10 advisories affecting 216 533 devices) accounted for 95% of device recalls. Implantable cardioverter-defibrillators were recalled more frequently than pacemakers (mean [SD], 16.4 [1.6] vs 6.7 [0.8] advisories per 100 person-years; P<.001). Between 1995 and 2000, the annual advisory rate increased for both pacemakers (P for trend <.001) and ICDs (P for trend =.02). An estimated 1.3 million device checks and analyses and 36 187 device replacements resulted from the advisories and cost approximately $870 million.
Conclusions:
Pacemaker and ICD recalls and safety alerts occur frequently, affect many patients, and appear to be increasing in number and rate. With the growing number of device implants and expanding indications for device therapy, the number of patients affected by device advisories will likely continue to increase.
More Related Videos
Related Concept Videos
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Instrument Transformers
Overcurrent Relays
Instantaneous overcurrent relays activate immediately when the input current exceeds a predetermined value, known as the pickup current, instantly energizing the circuit breaker trip coil. This rapid response is vital for addressing severe faults quickly.
Time-delay overcurrent relays, on the other...
Reclosers and Fuses
A comprehensive protection scheme for radial distribution...
Differential Relays
Cardiopulmonary Resuscitation III: AED Use