Geographical miss during catheter-based intracoronary beta-radiation: incidence and implications in the BRIE study.
1Department of Cardiology, Thoraxcenter, Erasmus Medical Center Rotterdam, The Netherlands.
Objectives:
We sought to determine the incidence and causes of geographical miss (GM) and to evaluate its impact on edge restenosis after intracoronary beta-radiation therapy.
Background:
Edge restenosis is a limitation of intracoronary beta-radiation therapy. Geographical miss is the situation in which the radiation source does not fully cover the injured segment and may lead to edge restenosis.
Methods:
We analyzed 175 vessels treated according to the Beta-Radiation In Europe (BRIE) study protocol. The effective irradiated segment (EIRS) and both edges were studied with quantitative coronary angiography. The edges of the EIRS that were injured constituted the GM edges. Restenosis was defined as diameter stenosis >50% at follow-up. Geographical miss was determined by simultaneous electrocardiographic-matched, side-by-side projection of the source and balloons deflated at the injury site, in identical angiographic projections surrounded by contrast.
Results:
Geographical miss affected 41.2% of the edges and increased edge restenosis significantly compared with non-GM edges (16.3% vs. 4.3%, respectively, p = 0.004). Restenosis was increased both in the proximal (p = 0.05) and distal (p = 0.02) GM edges compared with noninjured edges. Geographical miss associated with stent injury significantly increased edge restenosis (p = 0.006), whereas GM related to balloon injury did not significantly increase edge restenosis (p = 0.35). The restenosis in the EIRS was similar between vessels with and without GM (24.3% and 21.6%, respectively, p = 0.8).
Conclusions:
Geographical miss is strongly associated with restenosis at the edges of the EIRS. This effect is more prominent when caused by stenting. Geographical miss does not increase restenosis in the EIRS.
More Related Videos
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Cardiac Catheterization I: Pre-Procedure Overview