Merits and limitations of the mode switching rate stabilization pacing algorithms in the implantable cardioverter
1Department of Cardiology, Academic Hospital Maastricht, Maastricht, The Netherlands. b.dijkman@cardio.azm.nl
Background:
The 7250 Jewel AF Medtronic model of ICD is the first implantable device in which both therapies for atrial arrhythmias and pacing algorithms for atrial arrhythmia prevention are available. Feasibility of that extensive atrial arrhythmia management requires correct and synergic functioning of different algorithms to control arrhythmias.
Methods And Results:
The ability of the new pacing algorithms to stabilize the atrial rate following termination of treated atrial arrhythmias was evaluated in the marker channel registration of 600 spontaneously occurring episodes in 15 patients with the Jewel AF. All patients (55+/-15 years) had structural heart disease and documented atrial and ventricular arrhythmias. Dual chamber rate stabilization pacing was present in 245 (41 %) of episodes following arrhythmia termination and was a part of the mode switching operation during which pacing was provided in the dynamic DDI mode. This algorithm could function as the atrial rate stabilization pacing only when there was a slow spontaneous atrial rhythm or in presence of atrial premature beats conducted to the ventricles with a normal AV time. In case of atrial premature beats with delayed or absent conduction to the ventricles and in case of ventricular premature beats, the algorithm stabilized the ventricular rate. The rate stabilization pacing in DDI mode during sinus rhythm following atrial arrhythmia termination was often extended in time due to the device-based definition of arrhythmia termination. This was also the case in patients, in whom the DDD mode with true atrial rate stabilization algorithm was programmed.
Conclusions:
The rate stabilization algorithms in the Jewel AF applied after atrial arrhythmia termination provide pacing that is not based on the timing of atrial events. Only under certain circumstances the algorithm can function as atrial rate stabilization pacing. Adjustments in availability and functioning of the rate stabilization algorithms might be of benefit for the clinical performance of pacing as part of device therapy for atrial arrhythmias.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Mechanism of Cardiac Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias VI: Management of Dysrhythmias
Cardiomyopathy V: Interprofessional Care
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias