Faulty design resulting in temporary pacemaker failure.

B Kleinman1, M Baumann, C Andrus

  • 1Anesthesia Service, Edward Hines Jr. Veterans Hospital, Hines, IL 60141, USA. bkleinm@luc.edu

Chest
|August 15, 2001
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Clinical presentation, MR imaging and outcome in children with myelin oligodendrocyte glycoprotein antibody-negative acute disseminated encephalomyelitis.

European journal of paediatric neurology : EJPN : official journal of the European Paediatric Neurology Society·2025
Same author

Corrigendum to "Autoantibody status, neuroradiological and clinical findings in children with acute cerebellitis" [Eur. J. Paediatr. Neurol. 47 (2023) 118-130].

European journal of paediatric neurology : EJPN : official journal of the European Paediatric Neurology Society·2024
Same author

Frequency of an intrathecal IgM synthesis and MRZ reaction in children with MS.

European journal of paediatric neurology : EJPN : official journal of the European Paediatric Neurology Society·2024
Same author

Autoantibody status, neuroradiological and clinical findings in children with acute cerebellitis.

European journal of paediatric neurology : EJPN : official journal of the European Paediatric Neurology Society·2024
Same author

Pulse Parameters and Thresholds for (ir)Reversible Electroporation on Hepatocellular Carcinoma Cells <i>in Vitro</i>.

Technology in cancer research & treatment·2023
Same author

Time- and dose-dependent volume decreases in subcortical grey matter structures of glioma patients after radio(chemo)therapy.

Clinical and translational radiation oncology·2022
Same journal

The effect of a two-day stay at high altitude (2500 m) on right ventricular afterload and oxygen delivery in patients with pulmonary vascular disease A randomized controlled crossover trial.

Chest·2026
Same journal

A Comparative Study of Radiation Exposure in Conventional and Robotic Bronchoscopy.

Chest·2026
Same journal

Independent Prognostic Contributions of Anti-Ro52 and Anti-MDA5 in Autoimmune-Associated Interstitial Lung Disease.

Chest·2026
Same journal

Lung aeration and gas exchange in SGA or AGA infants with moderate-severe BPD: secondary analysis of the PATH-BPD study.

Chest·2026
Same journal

Lung Cancer Incidence and Mortality after Negative Low-Dose CT Screening Results.

Chest·2026
Same journal

Symptom prevalence and impact on lung cancer risk in the SUMMIT study.

Chest·2026
See all related articles

A temporary pacemaker failed after heart surgery due to a design flaw, not user error. This highlights how equipment design can escalate human mistakes into critical patient events.

Area of Science:

  • Cardiovascular Surgery
  • Biomedical Engineering
  • Critical Care Medicine

Background:

  • Post-operative care following aortic valve replacement and coronary artery bypass surgery requires vigilant monitoring.
  • Temporary pacemakers are crucial for managing cardiac rhythm disturbances in critically ill patients.

Observation:

  • A 77-year-old male patient experienced asystole three days after undergoing aortic valve replacement and coronary artery bypass surgery.
  • A dual-chamber temporary pacemaker generator failed to discharge upon activation, displaying an obscure error message.
  • Intensive care unit (ICU) staff were unable to resolve the pacemaker malfunction.

Findings:

  • The pacemaker failure was attributed to an inherent design flaw within the medical device.
  • The obscure error message hindered effective troubleshooting by clinical staff.

Related Experiment Videos

Implications:

  • Medical equipment design flaws can transform potential human errors into life-threatening situations for patients.
  • This case underscores the critical need for robust medical device design and rigorous pre-market testing.
  • Improved user interfaces and error reporting in medical devices are essential for patient safety in critical care settings.