Related Experiment Video
Updated: Jul 30, 2026

Calibrated Passive Sampling - Multi-plot Field Measurements of NH3 Emissions with a Combination of Dynamic Tube Method and Passive Samplers
Published on: March 21, 2016
Installation and calibration of Kayzero-assisted NAA in three Central European countries via a Copernicus project
F De Corte1, R van Sluijs, A Simonits
1Ghent University, Institute for Nuclear Sciences, Laboratory of Analytical Chemistry, Gent, Belgium. frans.decorte@rug.ac.be
Abstract:
An account is given of the installation and calibration of k0-based NAA--assisted by the DSM Kayzero/Solcoi software package--at the KFKI-AEKI, Budapest, the NPI, Rez and the IJS, Ljubljana. Not only the calibration of the Ge-detectors and the irradiation facilities are discussed, but also other important topics such as gamma-spectrometric hard- and software, QC/QA of the IRMM-530 Al-Au flux monitor and the upgrade of the Kayzero/Solcoi code. The work was performed in the framework of a European Copernicus JRP, coordinated by the Laboratory of Analytical Chemistry, Gent, with DSM Research, Geleen, as the industrial partner.