Related Experiment Video
Updated: Aug 6, 2026

Adjunctive Diode Laser Therapy and Probiotic Lactobacillus Therapy in the Treatment of Periodontitis and Peri-Implant Disease
Published on: May 9, 2022
Use of electrosurgery and lasers in the presence of dental implants
C W Wilcox1, T M Wilwerding, P Watson
1Department of Fixed Prosthodontics, Creighton University School of Dentistry, Omaha, Nebraska 68178, USA. ProsMan@aol.com
Abstract:
Use of electrosurgery or laser surgery in the presence of metallic implants has been implicated in generating heat-induced injury to peri-implant bone, with the subsequent loss of osseointegration. Studies involving lasers offer conflicting results, while in the case of the electrosurg, little research has been published supporting or refuting these claims. This study measured local heat effects created by use of a unipolar electrosurgical unit, a bipolar electrosurgical unit, and a neodymium:yttrium-aluminum-garnet (Nd:YAG) laser. Absolute temperature increase was measured during an in vitro uncovering surgical procedure performed with each unit. Analysis of variance for repeated measures was performed. Second, absolute temperature increase for each unit was compared with a theoretical clinical limit of a 10 degrees C increase. The findings suggest that use of the unipolar electrosurgical unit should be avoided, while judicious use of both the bipolar unit or the laser unit should produce temperature profiles well within clinical limits.