Waveform optimization for internal atrial defibrillation: effects of waveform rounding, phase duration, and voltage
B J Kidwai1, J D Allen, M T Harbinson
1Regional Medical Cardiology Center, Royal Victoria Hospital, Belfast, Ireland.
Rounding biphasic waveforms for internal atrial defibrillation maintains efficacy while reducing energy. The optimal phase duration for this double rounded waveform is 6-10 ms, enabling effective cardioversion with lower energy requirements.
Area of Science:
- Cardiovascular Electrophysiology
- Medical Device Engineering
Background:
- Atrial fibrillation (AF) necessitates effective defibrillation strategies.
- Conventional truncated exponential biphasic waveforms are used for internal atrial defibrillation.
- Optimizing waveform parameters can improve energy efficiency and reduce adverse effects.
Purpose of the Study:
- To compare the efficacy of internal atrial defibrillation using conventional biphasic waveforms with and without waveform rounding.
- To determine the optimal phase duration for a novel double rounded waveform.
- To assess the impact of waveform modifications on delivered energy and cardioversion success.
Main Methods:
- Internal atrial defibrillation was performed in anesthetized sheep using defibrillation catheters.
- Conventional biphasic waveforms were compared with single and double rounded waveforms.
- Waveform parameters, including phase duration and voltage, were systematically varied.
- Cardioversion success rates and delivered energy were recorded.
Main Results:
- Rounding both phases of the biphasic waveform (double rounded) significantly reduced peak voltages, currents, and delivered energy (>21%) without compromising cardioversion success.
- The efficacy of single and double rounded waveforms was comparable to the conventional waveform at specific voltage settings.
- Optimal efficacy for the double rounded waveform was observed at 6-10 ms phase duration, achieving 100% success at 10 ms with reduced delivered energy.
- Successful cardioversion correlated strongly with peak-to-trough voltage swing across different waveform types.
Conclusions:
- The double rounded waveform offers a promising alternative for internal atrial defibrillation, providing similar efficacy to conventional biphasic waveforms but with reduced energy delivery.
- Optimizing phase duration to 6-10 ms is crucial for maximizing the effectiveness of the double rounded waveform.
- Waveform rounding presents a viable strategy for enhancing energy efficiency in internal atrial defibrillation devices.
Related Concept Videos
Effective Value of a Periodic Waveform
The effective value of a periodic current represents the direct current (DC) that conveys the same average power to a resistor as the periodic current itself. This concept is crucial when assessing AC circuits. To determine the...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Dysrhythmias III: Characteristics of Dysrhythmias
Dysrhythmias VI: Management of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias