Validation of a diffusive sampler for NO2
A Hagenbjörk-Gustafsson1, R Lindahl, J O Levin
1National Institute for Working Life, Department of Chemistry, P.O. Box 7654, S-907 13 Umeå, Sweden.
Abstract:
A diffusive sampler for NO2, Willems badge, was validated in laboratory experiments and field tests. The collecting reagent for NO2 in the sampler is triethanolamine, and the analysis is based on a modified colorimetric method, the Saltzman method. The analysis was performed by a flow injection analysis (FIA) technique. The sampling rate for the sampler was determined to be 40.0 ml min-1. There was no effect of NO2 concentration or relative humidity on sampling rate, and the influence of sampling time was found to be small. The detection limit was 4 micrograms m-3 for a 24 h sample. The capacity is high enough to allow sampling of 150 micrograms m-3 for 7 days, which is twice the recommended Swedish short-term (24 h) guideline value as a 98-percentile over 6 months. In field tests, the sampler performed well, even at wind speeds higher than 2 m s-1, and at low temperatures. The overall uncertainty of the method was 24%. The sensitivity and capacity of the method also make it suitable for personal sampling for 2-8 h in working environments.
Related Concept Videos
Behavior of Gas Molecules: Molecular Diffusion, Mean Free Path, and Effusion
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Passive Diffusion: Overview and Kinetics
When administered orally, drugs establish a substantial concentration gradient between the gastrointestinal (GI) lumen and the bloodstream, expediting their diffusion into...