Related Experiment Video
Updated: Oct 4, 2026

High-Temperature and High-Pressure In situ Magic Angle Spinning Nuclear Magnetic Resonance Spectroscopy
Published on: October 9, 2020
Rotor synchronization of radiofrequency and gradient pulses in high-resolution magic angle spinning NMR
J M Wieruszeski1, G Montagne, G Chessari
1UMR 8525 CNRS, Institut de Biologie de Lille, Institut Pasteur de Lille, Université de Lille II, 1 rue du Professeur Calmette, 59021 Lille Cedex, France.
Abstract:
We have investigated the extent to which rotor synchronization of radiofrequency pulses leads to spectral improvement in high-resolution magic angle spinning NMR experiments. Several pulse sequences were tested, and the effect was found to be maximal in homonuclear TOCSY spectra. The physicochemical nature of the sample plays a role in the phenomenon, as rotor synchronization allows the refocusing of residual anisotropic interactions. However, even in a liquid sample the effects were visible. Radial inhomogeneities of the radiofrequency field were identified as an important source of the problem.
Related Concept Videos
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Atomic Nuclei: Magnetic Resonance
NMR Spectrometers: Resolution and Error Correction
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...
NMR Spectrometers: Overview