Does the cleaning technique influence the durability of the <9F flexible ureteroscope?
E M McDougall1, G Alberts, K J Deal
1Department of Urology, Vanderbilt University Medical Center, Nashville, Tennessee 37232-2765, USA. elspeth.mcdougall@mcmail.vanderbilt.edu
Background And Purpose:
The fragility of the <9F flexible ureteroscope limits its availability to general urology practice. The purpose of this study was to determine whether the technique used to clean the flexible ureteroscope or the number of persons handling the instrument during the cleaning process influenced endoscope breakage or deterioration during regular endourologic use.
Patients And Methods:
A new Olympus URF/P3 flexible 7.5F ureteroscope was used for each of two 30-day study periods during which a single surgeon used the endoscope for a variety of upper urinary tract procedures. During the first 30-day period (Group 1), the endoscope was leak-proof-pressure tested and cleaned by the endourology support team using the Steris 20 (peroxyacetic acid 35%) technique. During the second 30-day period (Group 2), the endoscope was leak-proof tested and cleaned only by the surgeon using the Cidex (glutaraldehyde 2.4%) technique. A record was kept for each ureteroscopic case to document the patient position, access technique, time the endoscope was in the urinary tract, instruments passed through the ureteroscope, and the maximum irrigant pressure used. In addition, a record was made of the number of broken fibers, the degree of flexion and deflexion of the endoscope, and the problems encountered with the endoscope during the case.
Results:
The two study groups were similar in terms of the total number of cases performed, the mean time the endoscope was in the urinary tract per case, the access approach used, and the use of the ureteral access sheath and ancillary equipment. In Group 2, the endoscope was used for a longer total time (618 minutes v 457 minutes), and access to a lower pole calix was more than twice as common as in Group 1. This may explain why more broken fibers were noted in the instrument used in Group 2 over the study period (eight v four broken fibers) than in Group 1. The only breakage occurred as a result of the surgeon accidentally activating the laser probe inside the working channel of the endoscope in Group 2.
Conclusion:
The technique and number of personnel involved in the maintenance and cleaning of the flexible ureteroscope does not have a significant effect on the durability and function of these instruments. It is the arduous demands of the endourologic procedure that influence the durability of these fragile endoscopes.
Insights
Flexible ureteroscope durability is not affected by cleaning methods or personnel. The demands of endourologic procedures, not maintenance, impact flexible ureteroscope longevity.
Area of Science:
- Urology
- Medical Device Engineering
Background:
- Flexible ureteroscopes are fragile, limiting their use in general urology.
- Investigating factors affecting flexible ureteroscope breakage is crucial for wider adoption.
Purpose of the Study:
- To evaluate if cleaning techniques or the number of personnel handling the flexible ureteroscope impacts its breakage or deterioration.
- To understand the influence of maintenance on flexible ureteroscope durability.
Main Methods:
- A new flexible ureteroscope was used over two 30-day periods.
- Group 1: Endoscope cleaned by support team using Steris 20 (peroxyacetic acid).
- Group 2: Endoscope cleaned by surgeon using Cidex (glutaraldehyde).
- Detailed records of usage, procedures, and endoscope condition were maintained.
Main Results:
- Both groups had similar case numbers and usage patterns.
- Group 2 experienced more broken fibers (eight vs. four) and longer total usage time.
- The sole breakage in Group 2 resulted from accidental laser probe activation.
Conclusions:
- Cleaning technique and personnel involved do not significantly affect flexible ureteroscope durability.
- The demands of endourologic procedures are the primary factor influencing flexible ureteroscope longevity.